Based on our patented Permeance Measurement Technology Platform, combined with Constant Temperature Fiber Technology and Digestion Colorimetric Integration Technology, WDet-5000 Ammonia Nitrogen Water Quality Online Analyzer can achieve high precision, low detection limit, high stability and low maintenance of automatic ammonia nitrogen online monitoring. The product fully meets the application of pollution sources, industrial production process, municipal wastewater, etc. It has passed the certificate of CPA and CCEP.
Features
Permeance Measurement Technology will not be interfered by chromaticity, suspended solids and bubbles
High reliability, strong ability to resist chromaticity, calcium and magnesium ions
High detection precision, low LLOQ (0.01mg/L), small drift for long-term operation
With one button self-test and self-diagnosis function
Highly integrated valve terminal, each valve can be maintained, disassembled and cleaned individually
Able to extend to all-day networking for real-time inspection of instrument running state
Equipped with alarm function, quality control function and reverse control function
Applications
Applied in environmental pollution source monitoring, industrial process water monitoring, municipal sewage processing, surface water monitoring, etc.
Specifications
Measuring method: Salicylic Acid Spectrophotometry
Range: 0.01~2/20/100mg/L (extensible)
Indication error: ±8%(20%F.S.) /±5%(50%F.S.) /±3%(80%F.S.)
Repeatability: ≤2%
Zero drift: ≤0.02mg/L
Span drift:<1%
LLOQ: 0.01mg/L
Reagent: standard reagent can make 400 samples
Memory effect: 20%F.S. ±0.3mg/L
80%F.S. ±0.2mg/L
Measurement interval: point measurement, interval measurement, or be started via serial port
Maintenance:<2 hours/month
Ambient temperature: 5℃~45℃
Communication interface: RS485/RS232/USB interface (optional)
1x4-20mA input/output (can be expanded to 2)
Power consumption: 200W(220±10%VAC 50Hz), without sampling pump
Note: test method refers to the latest standards of Environmental Certification