The AM-5105 H2S Analyzer adopts Ultraviolet Fluorescence Method to provide accurate and reliable low-concentration H2S detection of “ppb~ppm” level, which can be applied to atmospheric monitoring. The built-in low-temperature reformer can convert H2S under 315℃ without being affected by other gases. The sample gas first passes through a cleaner to remove SO2 and then converter will turn H2 S into SO2. At last, SO2 concentration can be measured under fluorescence.
Features
Use standard 19-inch 4U chassis
Color touch screen
Provide RS232/RS485 communication port
Support remote control
Adopt Ultraviolet Fluorescence Method with low detection limit
With independent SO2 and H2S analog output
Applications
Applied to atmospheric monitoring
Specifications
Measuring range: 0~500nmol/mol(ppb), 0~20umol/mol(ppm) (adjustable)
Zero noise: <1ppb (RMS)
Limit of detection: 2ppb
Zero drift: <1ppb/24h
Span drift: <5ppb/24h
Linearity: <1%F.S.
Response time:<120s (T90)
Indication error: ±2%F.S.
Sampling flow range: 800cm3/min±10%
Ambient temperature: 5℃~40℃
Power: 220V~240V, 50/60Hz, 400W
Weight: <15kg
Analog output: 10V, 5V, 1V, 0.1V (optional)
Dimension: 178mmx432mmx613mm
Communication: RS232, RS485